Day 5: SDS-PAGE and Electrophoretic Transfer
Assignments Due
- Plagiarism tutorials
and exercises must be completed ONLINE by 11:30
pm on Fri., 21-Sept-2012
- Begin a Materials and Methods section
that includes the procedures through Day 5; include appropriate References (cited
as in Protein
Expression & Purification).
Bring a TYPED draft to Day 6 lab for evaluation.
See Research
Paper
- Introduction and References (McMillan:
pp. 59-61, 78-81, 107-125, 126-160 (3rd ed.); 76-78, 104-107,
5-32, 149-153, 167-205 (4th ed.); 69-70, 131-136 (5th ed.))
- Calculations for SDS-PAGE sample preparation in
your lab notebook (due at the beginning of lab)
- Review the Introduction
to SDS-PAGE, Preparing gels, Preparing samples, Running gels from Experimental
Biosciences Resources
- Gel electrophoresis,
western blot, isoelectric focusing (Scopes 2nd ed. pp.
287-292; 3rd ed. pp. 295-307)
- OWL-Space
BIOC 311 Resources (reveiw before Monday discussion)
- [Electrophoresis Resources]
- ready_cell.pdf
- mini_transblot.pdf
Preparation
- Writing the Introduction
- Electrophoresis applications in characterization of proteins
Overview of Experiment
In this session a denaturing 10% polyacrylamide gel (SDS-PAGE)
is run to separate the proteins based on size. We use the discontinuous
SDS-PAGE system developed by Laemmli, UK (Cleavage
of structural proteins during the assembly of the head of
bacteriophage T4. Nature 227 (5259): 680-685, 1970);
a resolving or separating (LOWER) gel and a stacking (UPPER)
gel are cast with different pore sizes, pH, and ionic strength.
Additionally, different mobile ions are used in the gel and
running buffers. This buffer discontinuity concentrates large
sample volumes in the stacking gel and allows high resolution
in the separating gel.
One half of the gel is silver stained to visualize
the protein bands in order to evaluate the progress of the purification
procedures
-- Silver stain is NONSPECIFIC and gives
a qualitative picture of the efficacy of the
purification scheme.
The other half of the gel is electrophoretically
transferred onto a PVDF membrane that becomes our western blot --
in a later class we will develop the membrane with an antibody
generated against murine adenosine deaminase to identify
SPECIFICALLY which band corresponds to adenosine deaminase
and to estimate molecular mass (m).
Procedures
Gel Contents
- 1st half: Stain to evaluate protein purification
- NATIVE purification fractions
- RECOMBINANT purification fractions
- Standards
- 2nd half: Transfer to PVDF membrane for
western analysis
- NATIVE fraction(s)
- RECOMBINANT fraction(s)
- Standards
NOTE: use a maximum of FIVE wells
on this half of the gel
SDS-PAGE Sample Preparation
NOTE:
- Make up your samples so that you can load the SAME
volume of each of them onto the gel. Add
Tris/glycine/SDS (25 mM Tris, pH 8.3, 192 mM glycine, 0.1%
SDS) running buffer to make up the difference in volumes.
- The maximum volume you can load into each well is
15 µl;
the 5X-SDS gel-loading buffer MUST be included in this 15 µl.
- Prepare at least 60 µl of each
sample (enough for 4 loadings). Even though a smaller
volume will actually be loaded onto the gel, the transfers
and heating steps result in loss of liquid making manipulations
of less than 10 µl samples a challenge.
Also, you and your partners will need to load one or more of
each sample onto each gel.
- Using the concentration of protein in each sample (in mg/ml or µg/µl), determine the volume to load into each well of the gel:
For the crude sample you need to load 7.5-10 µg of
protein per well. Prepare the sample in 12 µl; dilute if
necessary with the Tris/glycine/SDS running buffer. The crude
samples with multiple proteins must have more protein loaded
since the total protein will be distributed in many bands.
For the size exclusion and the Q fractions,
you need to load 0.5 to 2 µg per well.
Prepare the sample in 12 µl; dilute if necessary with the Tris/
glycine/SDS running buffer.
*if your fractions are very dilute, use the maximum sample volume (12 µl).
- Add 5X-SDS gel-loading buffer to each sample to give 1X final
conc. and MIX WELL. This solution contains a dye (bromophenol
blue) to monitor the electrophoretic front and glycerol to
encourage the sample to settle to the bottom of the well during
loading.
β-mercaptoethanol is also included
in the tracking dye to enhance denaturation by disrupting disulfide
bonds between cysteines.
1X SDS gel-loading buffer:
50 mM Tris-Cl (pH 6.8)
2% SDS
2% β-mercaptoethanol
0.1% bromophenol blue
10% glycerol
- Heat the samples in a boiling water bath for 4 minutes.
- Pulse spin the samples to collect the liquid into the bottom of the vials.
- Put samples on ice until ready to load the gel.
- Mix sample by pipetting before loading the gel.
Plate Assembly
Precast polyacrylamide mini gels from Bio-Rad [record
the specific details about the type of gel you use in your
lab notebook]
are assembled as described below. These instructions are
adapted from the BioRad Ready Gel Cell Instruction Manual.
- Get ONE gel per team.
- Take pre-cast gel out of its pouch and remove the
comb from the top of the gel.
- Carefully pull the tape off the bottom of the gel (the
tape must be removed so the bottom of the gel is exposed
to running buffer).
- Rotate the cams of the clamping frame outward; remove the
electrode assembly.
- Place the pre-cast gel into the slot at the bottom of the
electrode assembly.
NOTE: make sure that the SHORT glass plate faces INWARD
toward the notches on the green U-shaped gaskets.
**since you will only be using ONE gel, you will need to
use the mini cell buffer dam**
- Press gel and buffer dam up against the gaskets.
- Transfer the electrode assembly and gels into the clamping
frame.
- Gently press down on the electrode assembly while closing
the two cam levers of the clamping frame.
NOTE: Gently pressing the top of the electrode assembly
forces the top of the short plate on each gel against the
rubber gasket; if you do not push down, the upper buffer
chamber may leak.
- Carefully lower the assembled unit into the tank.
- Pour approximately 125 ml of Tris/glycine/SDS running
buffer (25 mM Tris, pH 8.3, 192 mM glycine, 0.1%
(w/v) SDS) into the center of the inner core until the
liquid level is slightly above the top of the inner glass
plate. Observe the bottom of the inner core for leaks.
If no leaks are found, pour approximately 200 ml of buffer
into the tank. Check for bubbles at the point of contact
between the gel and the buffer. Remove bubbles if necessary
using a disposable transfer pipet.
- The wells are outlined in "black": use regular
small tips for the pipettors and slowly load
your samples into the wells of the gel.
Half of the gel will be stained for protein and the other
half will be transferred for western analysis. Each half
of the gel should contain standards: the
prestained standards (Precision Plus Protein Kaleidoscope
Standards, BioRad) are ready to load (DO NOT BOIL); load
5 µl.
Load 15 µl of each of your samples (remember to load
the SAME volume for ALL).
Be sure to keep
a record of the well position of each sample.
- In the EMPTY wells, load 15 µl of 1X-SDS gel-loading
buffer in Tris/glycine/SDS running buffer; this step ensures
even migration of the dye front as the gel runs.
- Place the lid on the unit and connect the electrodes so that the anode is at the bottom of the gel. (The color scheme can be followed, red to red, etc.)
PLEASE NOTE: Banana plug fittings are not to be turned or twisted. Only push on and pull off by grasping the plug (or the entire lid for the boxes) without turning.
- Running conditions: Tris-HCl SDS gels are run at 200 V constant voltage. A typical run will take approximately 30 minutes.
Setting the Power Supply
- Turn on the power supply; switch is located on the right side.
- Set the voltage.
- Press the RIGHT SELECT button until DISPLAY is lit to monitor the actual voltage (v) and current (ma) during the run.
- Press RUN.
- STOP the run when the dye front is close to the bottom of
the glass plates.
NOTE: While the gel is running, pay attention
to the current display; if the current drops substantially or
if the gels are running very slow and the dye front looks distorted,
stop the run and add buffer to the inner chamber. The buffer
level must be ABOVE the inner plate.
Warning: Be certain the power supply reads "OFF" before handling the electrode connections. Electrical shock hazards are always present when using laboratory power supplies. Do not contact buffer reservoirs or wiring while the power supply is activated.
- When the run is complete, disassemble the gel. Using a
razor blade (Bio-Rad key), carefully
cut off the stacking gel from the gel; carefully cut the gel
in half.
BIO-RAD Precision Plus Protein Kaleidoscope Standards
Color
|
Molecular Mass (m), kD
|
Blue
|
250
|
Purple
|
150
|
Blue
|
100
|
Pink
|
75
|
Blue
|
50
|
Green
|
37
|
Pink
|
25
|
Blue
|
20
|
Blue
|
15
|
Yellow
|
10
|
NOTE: These standards must be recorded in your lab notebook.
Silver Stain of SDS-PAGE Gels (can detect as little as 0.5
ng per band)
Adapted from:
Morrissey, J. H. (1981). Silver stain for proteins
in polyacrylamide gels: a modified procedure with enhanced
uniform sensitivity. Anal. Biochem.117.2: 307-310.
Silver stain is up to 100X more sensitive
for detection of protein than the Coomassie stain and will
be used to stain the SDS-PAGE gels. Take care not to
touch the gel with your fingers as the stain will develop the
deposits left there.
Solutions
40% methanol, 10% acetic acid
*5µg/ml dithiothreitol (DTT)
*0.1% AgNO3
*3% Na2CO3 w/ formaldehyde
Milli-Q H2O (RO H2O is NOT clean enough for the silver stain
procedure)
[*these solutions were prepared in Milli-Q H2O]
Procedure
- Place the SDS-PAGE gel in a container (plastic storage
boxes work well; these are in the lab closet) and cover
with 40% methanol, 10%
acetic acid solution. Shake on rocker for 10 minutes.
- Wash the SDS-PAGE gel with excess Milli-Q H2O
three times for 5-10 minutes each time, on the rocker.
- Pour off the water after last wash and add 5µg/ml
DTT solution to cover gel. Shake for 10 minutes and pour
off DTT solution.
- Add 0.1% AgNO3 solution and shake for 10 minutes.
CAUTION: WEAR GLOVES! The AgNO3 solution is toxic and
should not touch your skin.
- Rinse with Milli-Q H2O.
- Add 3% NaCO3 w/ formaldehyde and watch gel for desired
intensity. If the developer soultion in the tray turns yellow
or brown within one minute, pour off and apply fresh developer.
When desired intensity of gel bands is reached, rinse briefly
with Milli-Q H2O and photograph the results (each team
should bring a DIGITAL CAMERA to lab--iPhone/iPad
will NOT work): place gel on
light box; please clean light box with a Kimwipe after disposing
of gel.
Note: The incubation times above are acceptable for gels up
to 1.5 mm thick. Extend times by 5 minutes each for thicker
gels.
Western Blot: Electrophoretic Transfer of Proteins from SDS-PAGE
*REFERENCE: Towbin, H., Staehelin, T., and Gordon,
J. (1979). Electrophoretic transfer of proteins from polyacrylamide
gels to nitrocellulose sheets: Procedure and some applications. Proc.
Natl. Acad. Sci. USA 76 (9): 4350-4354.
{Look at the Bio-Rad Mini Trans-Blot Cell Instruction Manual for detailed directions and illustrations.}
- Obtain a polyvinylidene difluoride (PVDF) membrane
(Immun-Blot® PVDF
Membranes, Bio-Rad) with the same dimensions as the gel and mark
it in a corner with a ball point pen to identify the orientation
of the membrane with the gel. Do not handle the gel or
membrane with your bare hands--wear gloves! Do not allow the
membrane to bend or fold.
- Obtain 700 ml per buffer tank of transfer
buffer (10 mM CAPS, pH 11.0, 10 % methanol).
- Wet the membrane in methanol for a couple seconds and immediately
place in transfer buffer; be
sure not to let the membrane dry out after it is wetted with
methanol.
NOTE: the PVDF membranes are hydrophobic and do not
wet well in aqueous solutions; if you SKIP the Methanol step,
the transfer will be poor. If the membrane dries
out, re-wet with methanol before putting in aqueous solutions.
- Completely saturate the pre-cut filter paper and fiber pads
of the transfer unit by soaking them in transfer buffer. Avoid
trapping air bubbles in the fiber pads or filter paper during
the wetting process.
Assembly of the gel holder cassette
The following procedures should be performed while wearing gloves to prevent contamination of the membrane.
- Open the gel holder cassette by sliding and lifting the latch. Note that one panel of the cassette is clear and the other panel is tinted smoky gray. The clear panel is the anode (+) side and the gray panel is the cathode (-) side. For the electrode module, the cathode is the smoky gray electrode panel located in the center of the buffer tank. Always insert the gel cassette so that the gray plastic faces the gray plastic of the cathode electrode. The anode and cathode are identified by the black (cathode) and red (anode) disks on top of the electrode module.
- Place the opened gel holder in a shallow container (plastic
storage box) so that the gray panel is flat on the bottom of
the container. The clear panel should rest at an angle against
the wall of the vessel. Pour some transfer buffer into the
container (just enough to cover the panel).
- Place the fiber pad on the gray panel of the cassette panel and be sure all air is removed. When assembling the fiber pads, filter paper, gel, and membrane, center all components in the holder. If any of the material extends past the edge of the cassette, it will catch on the guide rails, causing a distortion of both the blot-gel contact and the transfer pattern.
- Place a piece of the saturated filter paper on top of the fiber pad. Be sure all air is removed from the filter paper and the layer below it. Place the gel on top of the paper. Align the gel in the center of the cassette. Transfer will be incomplete for any portion of the gel that has bubbles on the surface or lies outside the pattern of circles of the gel holder cassette. Make sure that no air bubbles are trapped between the gel and the filter paper. Bubbles can be removed by rolling a plastic test tube over the filter paper and gel.
- Place the prewetted PVDF membrane on top of the gel. This is best done by holding the membrane at opposite ends so that the center portion will contact the gel first. Then, gradually lower the ends. Next, roll a plastic test tube over the top of the membrane to exclude all air bubbles from the area between the gel and membrane.
- Complete the sandwich by placing another piece of saturated filter paper on top of the membrane and placing a saturated fiber pad on top of the filter paper.
- Close the cassette. Hold it firmly so the sandwich will not move and secure the latch. At this point, movement of the sandwich might disrupt the gel-membrane contact, which would cause incomplete transfers or swirling transfer patterns.
- Place the gel holder in the buffer tank so that the gray panel of the holder is facing the gray cathode electrode panel.
- Prepare a "dummy" cassette with two fiber pads
(takes up volume so less transfer buffer is needed).
- Put the cooling tray in the back of the chamber. (The cooling
tray containing "freezer pack" can be found in the chest freezer
in B06.)
- Add the transfer buffer from the plastic container to the tank.
- If necessary, add more transfer buffer to cover the gels. (Do NOT fill the tank all the way to the top.)
- Place a stir bar in the buffer tank and use a
stir motor to keep temperatures low during the transfer.
- Turn on the magnetic stirrer, and put the lid in place. Be sure that the electrode wires on the lid are attached to the appropriate pins of the electrode module. Plug the unit into the power supply. Normal transfer polarity is cathode to anode, i.e. red wire to red outlet and black wire to black outlet on the power supply.
- Turn on the power supply to initiate transfer. Set the voltage to 50V and transfer for 1 hour.
Warning: Be certain the power supply reads "OFF" before handling the electrode connections. Electrical shock hazards are always present when using laboratory power supplies. Do not contact buffer reservoirs or wiring while the power supply is activated.
- After transfer is complete, turn off the power supply and disconnect the electrodes. Disassemble the cassette and mark the side of the membrane that was in contact with the gel if it is not already marked.
- Make pen marks on the membrane beside the colored standards and record their location as a sketch in the notebook.
- Place the membrane inside a clean paper towel and store at room temperature in your drawer.
PLEASE NOTE: The exposed wires in the electrophoresis
equipment are platinum and are very expensive to replace. Please,
only rinse these components with RO water, do not use a
brush,
and set to air dry. Be careful handling the devices to avoid
breaking the wires. For drying, stack the units on the desk
top so that no component is inside the other.
Do not throw away the pre-cut fiber pads--they are $5 each!
Copyright, Acknowledgements,
and Intended Use
Created by B. Beason (bbeason@rice.edu), Rice University, 15 June 1999